A Flexible, Customized Platform
To Give You Just What You Need.
The Reveal platform gives you a customized, in-line solution along with
flexibility on how to operate and integrate it. You have choices with Reveal.
We can build the application modeling for you, or you can build your own.
POWER
AC line power 120-265 VAC, 50/60 Hz
DC Power 24V
Dual hot-swappable batteries
DETECTOR
UV (190-400 nm)
UV/VIS (200-850 nm)
VIS (350-850 nm)
NIR (1550-1950 nm)
Extended NIR (1100-2100 nm)
LIF (280-800 nm)
CONTROLLING INTERFACE/MODELING
NovaPAC/NovaMath – user defined models
Reveal – Internal model
EXTERNAL CONNECTIONS
Measurement interface, industrial
communications, power
CHASSIS OPTIONS
Quick-release brackets
ATEX
NFPA
Heating, cooling
DIMENSIONS
300 mm x 210 mm x 110 mm
WEIGHT
13 lbs.
CONSTRUCTION MATERIALS
Enclosure: 316 stainless, anodized aluminum
Sample Interface: 316 stainless, sapphire (std.)
Others available as required
MAINTENANCE
Automated calibration/referencing
User replaceable lamp cartridge
(1500+ operating hours)
PROCESS INTERFACES
100’s of reflectance probes,
transflectance probes, flow cells
LIGHT SOURCE
Tungsten halogen lamp
Deuterium lamp
Combo tungsten / deuterium lamp
LED
COMMUNICATIONS
Ethernet, WiFi, Modbus, OPC UA, EtherNet/IP,
expansion port for additional outs
It runs 24/7/365 in any environment without ongoing operator interaction. Our internal calibration hardware allows for stability and reliability.
All Reveal enclosures are IP65 or higher. Hazardous area options are also available.
Dimensional drawings in mm
Our fully-integrated software suite handles hardware control and chemometric analysis while ensuring that Reveal remains compliant with the U.S. FDA’s strictest guidelines on electronic records and signatures (21 CFR Part 11), the PAT initiative and Good Automated Manufacturing Practices (GAMP) as well as U.S. and European Pharmacopeia.
Having the right process interface is essential to making any measurement. We have hundreds of different probes that fit in almost any environment:
- Pressure: less than 11,000 psi
- Operating temperature of machine: 0°C to 40°C (in standard packaging)
- Operating temperature of probe: under 400°C
MEASURING HEAD
- Use for non-contact measurement through a window
- LIF, NIR
- For Powders
- For Slurries
- Diffuse Reflectance, Fluorescence
- Variable Working Distance
- IP65 Compliant
CONTACT PROBE
- Use for contact measurement (at the probe tip)
- LIF, NIR For Powders: Diffuse Reflectance, Fluorescence
- For Liquids: Fluorescence
- IP65 Compliant
DIP PROBE
- Use for contact measurement via insertion
- UV, VIS, NIR, LIF For Liquids: Trans Reflectance, Transmission
- Multiple Optical Path Lengths Available
- Multiple Probe Sizes Available
ATR PROBE
- Use for contact measurement via insertion
- Excellent for strongly absorbing liquids, viscous liquids
- UV, VIS, NIR, LIF
- For Liquids
- For Slurries
- 3-Bounce ATR Design
- Sapphire Crystal Tip
FLOW CELL
- Use for flowing liquids
- UV, VIS, NIR, LIF
- Small Volume to Large Volume Available
- Piping Connections are Full Customizable
- Multiple Optical Path Lengths, Variable Path Available
HEATED PROBE
- Use for contact measurement via insertion
- Excellent for sticky substances; HSWG / Drying
- NIR
- For Powders: Diffuse Reflectance
- Sapphire Window
- Active Control: Heating
PURGED PROBE
- Use for contact measurement via insertion
- Excellent for Drying
- For Powders: Diffuse Reflectance
- Active Control: Compressed Air for Purging, Cleaning
It offers a full line of industrial communications options for easy integration into your control system, including Modbus, OPC and EtherNet/IP. An expansion port allows additional communication options, which is helpful if you want analog outs. There’s also a secured easy access to each Reveal so plant operators with tablets or smartphones can access the instrument in the process line.